Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-37
NGA2000 Reference
MINRANGE
Direction
Analyzer: bidirectional; 1/0 module: input
Binding
MINRANGE
Type
Floating point
Value range
0 - MAXRANGE
Units ppm
Description
The minimum range that the analyzer physics is capable of measuring.
Usage
This value is set in the manufacturing data screens, and is used to limit the range high
end selection in the expert set up screens. This value should only be changed if the
physics is changed so as to support the new value. It is determined by the noise level
of the physics for a given response time.
This is an input to the 1/0 module for informational purposes only.
See also
MAXRANGE
RNGHI
MOD_TEST_INS
Direction Output
Binding
None
Type Enumerated
Value range
Units None
Description
NDIR only; messages and instructions for the modulation test screens.
Usage
Used to show the present status of the modulation test, and to give instructions as to
what the user should do next. The MOD_TEST_xxx variables share data, they will
always contain the same value, although the phrases to which this value refers are
different for each variable.
See also
MOD_TEST_THN
MOD_TEST_TXT
MCHECKTIME
MOD_TEST_THN
Direction Output
Binding
None
Type Enumerated
Value range
Units None
Description
NDIR only; messages and instructions for the modulation test screens.
Usage
Used to show the present status of the modulation test, and to give instructions as to
what the user should do next. The MOD_TEXT_xxx variables share data, they will
always contain the same value, although the phrases to which this value refers are
different for each variable.
See also
MOD_TEST_INS
MOD_TEST_TXT
MCHECKTIME