Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-2
NGA2000 Reference
AM_INPUT
Direction Input
Binding
None
Type Enumerated
Value range
Analyzer specific: FID - LIGHT_FLAME; CLD- NOX_MODE, OZONATOR_ON; Not
present in NDIR or PMD modules.
Units None
Description
Shows which functions can be controlled by an 1/0 Module's digital input
Usage
Used to select the operation of the 1/0 Module's digital input line. Its value appears
on the 1/0 Modules' menus that control this function.
See also
ST1CONT
AMBC
Direction Bidirectional
Binding
AMBC
Type String
Value range
31 characters
Units None
Description
Describes the analyzer physical configuration (e.g. bench type).
Usage
This is a descriptive variable. It should not be changed unless the analyzer hardware
is changed.
See also
TAG
AMSN
GAS
AMDELAYTIME
Direction Bidirectional
Binding
None
Type Float
Value range
0-30
Units Seconds
Description
Delays the primary reading by this amount
Usage
Used to delay the output of one analyzer with respect to another, so as to cause their
responses to a gas change to be aligned in time. This is only useful for systems of
analyzers.
See also
AFT90_