Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-62
NGA2000 Reference
ZERO_GOOD
Direction Bidirectional
Binding
None
Type
4-element floating point array
Value range
Around 500,000 except CLD, 32,000
Units None
Description
User stored known good value of zero offsets.
Usage
Once a known good zero calibration has been achieved, the values of the zero offsets
involved may be stored by the user in this array. Another function will automatically
copy them back to the operational values (ZERO) if desired.
See also
SPAN_GOOD
ZERO
ZERO_WAS
ZERO_WAS
Direction Bidirectional
Binding
None
Type
4-element floating point array
Value range
Around 500,000 except CLD, 32,000
Units None
Description
Manufacturing checkout time stored known good value of zero offsets.
Usage
Once a known good zero calibration has been achieved, the values of the zero offsets
involved are stored by manufacturing check out in this array. There is a function that
will automatically copy them back to the operational values (ZERO) if desired.
See also
SPANWAS
ZERO
ZERO_GOOD
ZEROGAS
Direction Bidirectional
Binding
None
Type
4-element floating point array
Value range
Normally 0
Units ppm
Description
Zero gas. Element 1 refers to the zero gas for range 1, etc.
Usage
During zero calibration, the elements of ZERO are adjusted such that the value of PVA
resulting is within 0.02% of the full scale range value of the appropriate element of
ZEROGAS. If CALCHCKLIMITS is set to enabled, this will fail if the change required is
greater than the percentage of the full scale value given in CALFAIL. The value of
ZERO should be the same as the value of RAW_SIGNAL while zero gas is flowing,
and a zero may be effected by simply copying the value of RAW_SIGNAL into the
appropriate element of ZERO.
See also
SPANGAS
ZERO