Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-16
NGA2000 Reference
CHOP_SPEED
Direction Output
Binding
None
Type
Floating point variable
Value range
5 ± 0.1
Units Hz
Description
NDIR only; Chopper speed.
Usage
The rotation rate of the NDIR chopper motor. This must be 5 Hz ± 0.1 Hz. Anything
else indicates a physical problem.
See also
CONTROL
Direction
Analyzer: bidirectional; 1/0 Module: input
Binding
CONTROL
Type Enumerated
Value range
LOCAL; REMOTE; AUTO
Units None
Description
Determines range selection source.
Usage
Used in the analyzer to bind to the same variable in the I/O modules, so that the
latter may be controlled through the analyzer module's menus. LOCAL - only the
control module (or PC interface) may change the range; REMOTE - the 1/0 module's
digital inputs may change the range, though this can be overridden by the control
module; AUTO - the 1/0 module's automatic range change algorithm is enabled. In
this case if the user changes the range, the autorange change algorithm may well
immediately change it back.
See also
CRANGE
CONV_BIAS
Direction Bidirectional
Binding
None
Type Floating
point
Value range
50
Units
°
C
Description
CLD only; bias temperature for the converter temperature control algorithm.
Usage
Used in the converter temperature control PID algorithm. This value should not be
adjusted.
See also
CONV_IGAIN
CONV_PGAIN
CONV_SETP
CONV_IGAIN
Direction Bidirectional
Binding
None
Type Floating
point
Value range
0.003
Units None
Description
CLD only; the integral "gain" for the converter temperature control algorithm.
Usage
Used in the control algorithm. This value has been experimentally determined, and
should not under any circumstances be changed.
See also
CONV_BIAS
CONV_PGAIN
CONV_SETP