Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management I/O Modules Variables
3-11
NGA2000 Reference
CALSTATB
Direction Input
Binding
CALSTAT
Type
Enumerated type
Value range
READY; ZEROING - WAIT; SPANNING - WAIT,
Units None
Description
SCAL only; shows present calibration status of associated analyzer.
Usage
Bound to the CALSTAT of analyzer 2, this shows analyzer 2's current calibration
status, and is used by the SCAL module as a signal as to what part of its sequence
to do next. For example, it updates AMFNO with a value meaning "Do a zero", and
then monitors CALSTATA(B, C or D) to see if the analyzer has both received the
message and is doing a zero. Once it has done, and CALSTAT is back to normal
again, it goes on to the span step. If the analyzer does not update CALSTAT with
its zeroing message, the SCAL module realizes that the analyzer zero has failed,
and it will abort the calibration cycle at this point.
See also
CALSTATA
CALSTATB
CALSTATC
CALSTATD
CALSTATC
Direction Input
Binding
CALSTAT
Type Enumerated
type
Value range
READY; ZEROING - WAIT; SPANNING - WAIT.
Units None
Description
SCAL only; shows present calibration status of associated analyzer.
Usage
Bound to the CALSTAT of analyzer 3, this shows analyzer 3's current calibration
status, and is used by the S CAL module as a signal as to what part of its
sequence to do next. For example, it updates AMFNO with a value meaning 'Do a
zero", and then monitors CALSTATA(B, C or D) to see if the analyzer has both
received the message and is doing a zero. Once it has done, and CALSTAT is
back to normal again, it goes on to the span step. If the analyzer does not update
CALSTAT with its zeroing message, the SCAL module realizes that the analyzer
zero has failed, and it will abort the calibration cycle at this point.
See also
CALSTATA
CALSTATB
CALSTATC
CALSTATD