Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management I/O Modules Variables
3-17
NGA2000 Reference
IOVOLTS
Direction Output
Binding
None
Type
3-element floating point array.
Value range
Nominal power supply values, plus or minus a volt or so.
Units Volts
Description
The present value of the +15 V power supply.
The present value of the -15 V power supply.
The present value of the +5 V power supply.
Usage
The present power supply voltages. Of course you will be lucky to read this
variable if the third element is significantly different from 5V as it would stop the
Neuron chip from working: if it is different it probably indicates that the ADC or its
reference voltage is faulty.
See also
IOVOLTSWAS
IOVOLTSWAS
Direction output
Binding
None
Type
3-element floating point array.
Value range
Nominal power supply values, plus or minus a volt or so.
Units Volts
Description
Stored values of the power supply voltage array IOVOLTS.
Usage
Stores the values of the power supply voltages at manufacturing check out time.
Compare the present values to this in order to see what has changed. Only write
into this if the electronic hardware has been changed. There is a special function
in the manufacturing data screen which automatically stores this and other such
variables.
See also
IOVOLTS
K0TAG
Direction Output
Binding
None
Type String
Value range
29 characters
Units None
Description
SCAL only; module tag (as distinct from the subnode tags)
Usage
The tag for the base subnode. This has to be present for networking reasons.
See also
K1TAG
K1TAG
K1TAG, K2TAG, K3TAG, and K4TAG
Direction Output
Binding
None
Type String
Value range
31 characters
Units None
Description
SCAL only; subnode tags (as distinct from the module base tag)
Usage
The tags that select subnodes in the SCAL module. These are used to select
which subnode to bind to which analyzer. The tags are generated automatically
from the base tag in SUBNODE_BASE.
See also
K0TAG
TAG
SUBNODE_BASE