Reference Manual
748384-C
September 2003
Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-56
NGA2000 Reference
SVNAME
Direction
Analyzer and 1/0 module: output
Binding
None
Type
4-element enumerated array
Value range
A list of variable descriptions, each element having the same list.
Units None
Description
Contains the descriptions of the variables that the control module will use in the
secondary variable display, as well as the means of selecting the variables in the first
place.
Usage
The variable SVCONT shares data with this variable. When a selection is made in
SVNAME, it causes the control module to use the contents of the equivalent
enumeration in SVCONT as the value to be displayed on the auxiliary line of the single
analyzer display screen. Element 1 is displayed on line 1, and so on. The description in
SVNAME is used as the text description on the variable line on the same screen.
See also
SVCONT
SW_DIAG1A
SW_DIAG1B, SW_DIAG1C, SW_DIAG1D, SW_DIAG2A, SW_DIAG2B, SW_DIAG2C, SW_DIAG2D
Direction Output
Binding
None
Type String
Value range
31 characters
Units None
Description
Software error reporting variables.
Usage
Contain a series of strings reporting any software error detected by the internal
diagnostic function. Only the first error in a series will be reported. The values may be
reset to blank by editing SW_RESET to 1 (Reset).
See also
SW_RESET
SW_RESET
Direction Bidirectional
Binding
None
Type Enumerated
Value range
Report; Reset
Units None
Description
Resets the software error reporting variables.
Usage
If an error has occurred, the software error reporting variables will contain the error.
They can be reset by editing this variable to the value 1, or "Reset".
See also
SW_DIAG1A etc.
TAG
Direction
Analyzer: bidirectional; 1/0 module: output
Binding
ANALTAG
Type String
Value range
31 characters
Units None
Description
Analyzer, AIO, and ACAL: user tag for the module.
Usage
This is the tag for the module. It appears on the top line of every menu applying to that
analyzer. It appears as the selection on all the menus that allow the user to enter the
1/0 module menus. It is also used to select the module during the binding process. It
may be changed as desired via the menus.
See also
AMSN
AMTAG
KOTAG