Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management I/O Modules Variables
3-15
NGA2000 Reference
GENPARMC
Direction output
Binding
None
Type Enumerated
Value range
0; 0.1; 0.5; 1; 2; 5; 10; 30
Units Seconds
Description
AIO only; determines the delay before activation of any concentration alarm.
Usage
Setting this variable to a value other than zero will cause a delay in the activation
of a concentration alarm. This avoids spurious alarms in a noisy system where
short term spikes do not indicate an alarm situation. This my be particularly useful
in cases of extreme electrical interference, where spurious glitches cannot be
prevented from upsetting the measurement.
See also
ALLVL
RNGHI
ILINE
Direction Output
Binding
None
Type
7-element enumerated array
Value range
OFF; ON
Units None
Description
Shows the present state of the digital input lines to the 1/0 module.
Elements 1 through 6 show the input lines 1 through 6.
Element 7 shows the presence of the "GM" jumper, a jumper that connects the
output relays in a way that
Usage
This is used primarily as a diagnostic indication. It shows the presence or absence
of a voltage on the digital input lines. It may also be used to read the state of these
lines for general purpose uses, by making a PC poll this variable. In this case the
1/0 module acts as an input port for the LON.
The lines are used for the following functions:
Initiates a zero calibration in both ACAL and SCAL
Initiates a span operation both ACAL and SCAL
Initiates a complete zero and span calibration in both ACAL and SCAL
The LSB for the range to be calibrated on ACAL, it selects local or remote
calibration on the SCAL
The MSB for the range to be calibrated on the ACAL only, it. has no function on
the SCAL.
Has no function.
See also
ILINE_CNTRL
ILINE_CNTRL
Direction Output
Binding
None
Type Enumerated
Value range
DISABLED; ENABLED
Units None
Description
ACAL, SCAL only; determines whether the digital input lines can control the
calibration function.
Usage
If enabled, this allows the digital input lines to the 1/0 module to control the
calibration function.
See also
ILINE