Reference Manual
748384-C
September 2003
I/O Modules Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
3-22
NGA2000 Reference
RMEASURE
Direction output
Binding
None
Type
7-element enumerated array
Value range
OFF; ON
Units None
Description
Shows the present state of each relay power.
Usage
Each element shows whether the associated relay is powered or not. This is used
for diagnostic purposes to detect frozen relays.
See also
ILINE
RSTATUS
RNG1EI
Direction output
Binding
None
Type Enumerated
Value range
0-5V; 0-20mA; 4-2OmA.
Units None
Description
AIO, ACAL only; selects the analog output mode.
Usage
The analog output circuit produces both voltage and current at once, but only one
has meaning. The output is scaled according to the setting of this variable. If it is
set to 0-5V, the voltage output will cover a range of 0-5V when the input varies
between RNGLO and RNGHI on that range. This output is calibrated with the
variable CAL-, and the algorithm uses the appropriate elements of CAL
depending on the setting of RNG1EI.
See also
PPU
PTU
RNGPAR
Direction output
Binding
None
Type
4-element integer array
Value range
0 -100 except element 4, 0 - 300
Units None
Description
AIO, ACAL only.
The percent of range for the auto-range change algorithm to change down.
The percent of range to change up.
The initial range change time delay.
The range change toggle hold-off time delay.
Usage
Used to control the automatic range change algorithm. Elements 1 and 2 refer to
percentages of the next range down. Element 3 delays the range change so as to
avoid nuisance changes on short term spikes, and element 4 is further insurance
against range toggling.
See also