Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-29
NGA2000 Reference
HIST2V_
Direction Output
Binding
None
Type
2-element floating point array
Value range
Same as PRESS_IS for CLD
Units hPa
Description
CLD only; values of PRESS-IS at manufacturing check out time
Usage
Provides historical data on how the analyzer was set up during final check out
See also
PRESS_IS
I_TIME
Direction
Analyzer: bidirectional; I/0 module: input
Binding
TIME
Type
7-element integer array
Value range
Appropriate date and time values
Units None
Description
Periodic time update variable
Year
Mouth
Day
Hour
Minute
Second
Day of week -not implemented
Usage
Used to synchronize the internal clocks of analyzers and 1/0 modules with that of the
control module. The CM updates this variable every few seconds, and if it is updated
twice within 5 seconds, the analyzer or 1/0 module updates its internal clock to this
time.
See also
IGNITECYCLES
Direction Bidirectional
Binding
None
Type Integer
Value range
1-5
Units None
Description
FID only; the number of ignition attempts allowed
Usage
Limits the number of ignition attempts during an automatic ignition cycle. If the flame
has not ignited by this time, the algorithm will give up.
See also
IGNOVERIDE
Direction Bidirectional
Binding
None
Type Integer
Value range
2-10
Units none
Description
FID only; the duration of the glow plug current flow
Usage
The glow plug requires a current flow for a certain period in order for it to warm up
enough to ignite the fuel. This variable sets that time. It is normally 3 seconds
See also
IGNITECYCLES