Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-3
NGA2000 Reference
AMFN
Direction Bidirectional
Binding
AMFN
Type Integer
Value range
0 No operation
I Zero on the gas currently present
2 Span on the gas currently present
4 FID light flame; WNOX dump drain valve
8 Analyzer specific function
16 Refresh all network variables over the network
32 Measure linearization gas
64 Calculate fourth order polynomial from stored gas data
127 Abort present operation
128 Measure linearization "tweak" gas
More may be added in the future.
Units None
Description
Initiates and aborts specific analyzer functions.
Usage
Set this to 0 before and after any other number - the analyzer responds to a change
in this variable's value. Leave it on the desired value for a second before changing it
back to 0. This variable is used by other modules on the network to initiate actions in
the analyzer, and may also be used by PC programs to do the same thing.
See also
ACFN in autocal modules
AMHR
Direction Bidirectional
Binding
AMHR
Type String
Value range
31 characters
Units None
Description
Describes the hardware revision level
Usage
Descriptive only, this is not currently in use. AMBC is usually used instead.
See also
AMSN
TAG
AMBC
AMLSDATE
Direction Bidirectional
Binding
None
Type String
Value range
31 characters
Units None
Description
The last service date
Usage
Automatically loaded with the current date and time when the menus are entered for
this analyzer.
See also
AMSERVDATE
AMMFGDATE