Reference Manual
748384-C
September 2003
Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-28
NGA2000 Reference
HEATER_IGAIN
Direction Bidirectional
Binding
None
Type Floating
point
Value range
0.003
Units None
Description
NDIR, PMD only; the integral "gain" for the temperature control algorithm
Usage
Used in the control algorithm. This value has been experimentally determined, and
should not under any circumstances be changed
See also
HEATER_BIAS
HEATER_PGAIN
HEATER_SETP
HEATER_PGAIN
Direction Bidirectional
Binding
None
Type Floating
point
Value range
0.2
Units None
Description
NDIR, PMD only; the proportional gain for the temperature control algorithm
Usage
Used for the control algorithm. This value has been experimentally determined, and
should not under any circumstances be changed.
See also
HEATER_BIAS
HEATER_IGAIN
HEATER_SETP
HEATER_SETP
Direction Bidirectional
Binding
None
Type Floating
point
Value range
0-50
Units
°
C
Description
NDIR, PMD only; the set point for the temperature control algorithm.
Usage
Used to determine the set point for the control algorithm. This value has been
experimentally determined, and should not under any circumstances be changed.
See also
HEATER_IGAIN
HEATER_PGAIN
HEATER_BIAS
HIST2T_
Direction Output
Binding
None
Type
3-element floating point array
Value range
Same as TEMP_IS for CLD
Units
°
C
Description
CLD only; values of TEMP_IS at manufacturing check out time
Usage
Provides historical data on how the analyzer was set up during final check out
See also
TEMP_IS