Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-48
NGA2000 Reference
PWASA
Direction Output
Binding
None
Type Integer
Value range
Between 0 and 100
Units None
Description
NDIR only; Signal gain setting at manufacturing time.
Usage
Historical values of SIGNAL-GAIN. This show how the instrument was working when
it was tested during the manufacturing process. This value is loaded automatically by
the manufacturing data storage function.
See also
SIGNAL_GAIN
PWM_CUR_IS
Direction Output
Binding
None
Type
2-element floating point array except FID, 3-elements
Value range
Both elements between 0 and 5000.
Units mA
Description
Heater Current (NDIR, FID), Block heater current (CLD)
Fan Current (NDIR, FID), Converter heater current (CLD)
Air FET Current (FID)
Usage
Shows the current through the control FET for these devices. Used as a diagnostic
tool.
See also
PWM_DUTY
Direction Output
Binding
None
Type
NDIR, PMD, HFID: 2-element floating point array; FU): single floating point; CLD:
3-element floating point.
Value range
Both elements between 0 and 100
Units None
Description
Fan FET duty cycle (NDIR, PMD), Temperature controller duty cycle (FID, HFID), Fan
duty cycle (CLD).
Heater FET duty cycle (NDIR, PMD), Oven controller duty cycle (HFID), Converter
duty cycle (CLD).
Detector cooler (Peltier) duty cycle (CLD).
Usage
Shows the duty cycle of the control algorithm for these devices.
See also