Reference Manual
748384-C
September 2003
Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-60
NGA2000 Reference
TRENDVAR
Direction Bidirectional
Binding
None
Type
2-element enumerated array.
Value range
A list of analyzer variable descriptions
Units None
Description
Lists possible analyzer variables whose 15 minute averages may be stored in the
internal storage buffer.
Usage
This variable shares data with TRENDCONT. TRENDVAR contains enumerations
whose phrases describe the variables whose names are referred to by the
enumerations in TRENDCONT. In the menu system, the user selects an element in
TRENDVAR and the system tells the software in the analyzer to store the data in the
variable pointed to by TRENCONT.
See also
TRENDCONT
TRENDTIMEOUT
Direction Bidirectional
Binding
None
Type Enumerated
Value range
DISABLED; ENABLED
Units None
Description
Disables the control module time out function when the analyzer trend display is on the
screen. Not used.
Usage
Not used in version 2.2 or earlier, as graphic screens are not implemented in these
versions.
See also
TRENDVAR
TWEAK
Direction Bidirectional
Binding
None
Type
4-element enumerated array
Value range
DISABLED; ENABLED
Units None
Description
Determines whether the tweak correction is enabled on each range.
Usage
When enabled, the "Tweak' function will add additional linearization to the signal so as
to force the reading to go through the mid point gas values exactly.
See also
MIDPOINTA etc.
MID_GASA etc.
VALIDITY
Direction
Analyzer: output; 1/0 module: input
Binding
VALIDITY
Type Enumerated
Value range
VALID; INVALID; NOT FROM PROCESS
Units None
Description
Describes the validity of the PVA variable.
Usage
Reports the PVA invalid if any of the following are true:
Flow below limits
Invalid calibration on this range
Flame out (FID only)
Ozonator switch off (CLD only)
PROCESS variable reported anything but 0 (i.e. not from process).
See also
OPSTATUS
GENERALSTATE
PVA