Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-5
NGA2000 Reference
AMSN
Direction Bidirectional
Binding
AMSN
Type String
Value range
31 characters
Units None
Description
The module serial number.
Usage
Set by the manufacturer when the module passes final test during the manufacturing
process. This should not be changed lightly! It should match the serial number on the
analyzer case. If the analyzer computer board is exchanged, this value along with all
others should be corrected.
See also
AMHR
AMBC
TAG
AMSR
Direction Bidirectional
Binding
AMSR
Type String
Value range
31 characters
Units None
Description
The module software revision number.
Usage
Set by the software compilation process. This variable may be written to, but it is not
stored in non-volatile memory, so its value will revert to the ROM value on a power
cycle. In general, you can only use analyzers of a given revision level with control
modules of the same or higher level revision, at least up to revision 3.0..
See also
AMHR
AMBC
TAG
AMVOLTSWAS
Direction
Output only
Binding
None
Type
4 to 6-element floating point array. See VOLTS-IS for details.
Value range
Nominal power supply values, plus or minus a volt or so.
Units Volts
Description
Stored values of the power supply voltage array VOLTS-IS.
Usage
Stores the values of the power supply voltages at manufacturing check out time.
Compare the present values to this in order to see what has changed. Only write into
this if the electronic hardware has been changed. There is a special function in the
manufacturing data screen which automatically stores this and other such variables.
See also
VOLTS_IS
PRES_WAS
FLOW_WAS
TEMP_WAS
ZERO_WAS
SPANWAS
PVWAS