Reference Manual
748384-C
September 2003
Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-8
NGA2000 Reference
BLOK_IGAIN
Direction Bidirectional
Binding
None
Type Floating
point
Value range
0.003
Units None
Description
CLD only; the integral "gain" for the sensor block temperature control algorithm.
Usage
Used in the control algorithm. This value has been experimentally determined, and
should not under any circumstances be changed.
See also
BLOK_BIAS
BLOK_PGAIN
BLOK_SETP
BLOK_PGAIN
Direction Bidirectional
Binding
None
Type Floating
point
Value range
0.2
Units None
Description
CLD only; the proportional gain for the sensor block temperature control algorithm.
Usage
Used for the control algorithm. This value has been experimentally determined, and
should not under any circumstances be changed.
See also
BLOK_BIAS
BLOK_IGAIN
BLOK_SETP
BLOK_SETP
Direction Bidirectional
Binding
None
Type Floating
point
Value range
0-50
Units Degrees
C
Description
CLD only; the set point for the sensor block temperature control algorithm.
Usage
Used to ' determine the set point for the control algorithm. This value has been
experimentally determined, and should not under any circumstances be changed.
See also
BLOK_IGAIN
BLOK_PGAIN
BLOK_BIAS
CAL_ERR_MSG
Direction Output
Binding
CAL_ERR_MSG
Type
2-element enumerated array
Value range
CAL OK: USER ABORTED; ZERO TIMED OUT; SPAN TIMED OUT; ZERO OVER
LIMIT; SPAN OVER LIMIT.
Units None
Description
Results of last zero (element 1) or span (element 2) operation
Usage
Shows the results of a zero or span operation. If the value is greater than 0, some
kind of failure occurred. ZERO (or SPAN) OVER I= means that the reading just
before the zero or span operation was in error by an amount greater than that in
CALFPC.
See also
CAL_VALIDITY
CALFPC