Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-55
NGA2000 Reference
STCONT
Direction
Analyzer output and 1/0 module output
Binding
STCONT
Type
8-element integer array
Value range
Bit field variables; odd elements are 16 bits, even elements use only the last 4 bits
Units None
Description
Each bit of the first element shows the state of the corresponding alarm named in
ST1NAME; the odd elements show the first 16 elements of ST1NAME, the even show
the last four. Similarly for the remaining pairs with ST2NAME etc.
Usage
Used to show the state of the specific analyzer alarms listed in ST1NAME, ST2NAME
and so on.
See also
ST1NAME
SVCONT
Direction
Analyzer and 1/0 module: output
Binding
None
Type
4-element enumerated array
Value range
A list of variable names, each element having the same list.
Units None
Description
Contains the variable names for the control module to use in the secondary variable
display.
Usage
The variable SVNAME shares data with this variable. When a selection is made in
SVNAME, it causes the control module to use the contents of the equivalent
enumeration in SVCONT as the value to be displayed on the auxiliary line of the single
analyzer display screen. Element 1 is displayed on line 1, and so on. The description in
SVNAME is used as the text description on the variable line on the same screen.
See also
SVNAME
SVFLL
Direction Bidirectional
Binding
None
Type
Floating point
Value range
0 - about 200
Units cc/min
Description
The sample bypass flow rate lower limit
Usage
Used in the alarm system to provide the lower limit for the flow alarm. When the flow-
rate given in FLOW_IS drops below this value, the variable VALIDITY is set to invalid.
See also
SVFUL
FLOW_IS
SVFUL
Direction Bidirectional
Binding
None
Type
Floating point
Value range
600-1200
Units cc/min
Description
The sample bypass flow rate upper limit
Usage
Used in the alarm system to provide the upper limit for the flow alarm.
See also
SVFLL
FLOW_IS