Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-31
NGA2000 Reference
LASTZERO
Direction Output
Binding
None
Type Floating
point
Value range
0 - the analyzer PVA Emit
Units ppm
Description
The last zero gas reading, using the SPAN factor from the subsequent span
operation.
Usage
Shows any interaction between zero and span operations. If the span operation
affects the zero, this will show a change. It is most useful when trying to calibrate an
analyzer with a non-zero zero gas.
See also
LASTSPAN
LIN_CASETEMP
Direction Bidirectional
Binding
None
Type
Floating point
Value range
0-70
Units
°
C
Description
The temperature at which the linearizer curve was made.
Usage
Used for temperature compensation of analyzers. The signal should be adjusted for
what it would be at the temperature that the analyzer was linearized, as the
linearization curve is itself dependent upon temperature. This variable contains that
value.
See also
Other LINxxxxx variables
LIN_ERROR
Direction Output
Binding
None
Type
Floating point
Value range
0-100
Units Percent
Description
Not used.
Usage Not
used.
See also
LIN_GASESA
LIN_GASESB, LIN_GASESC
Direction Bidirectional
Binding
LIN_GASESA, etc.
Type
7-element floating point (except LIN_GASESC, 6-element) array
Value range
-1 - 1,000,000 (-1 shows that no value has been entered)
Units ppm
Description
Linearization function actual gas input value.
Usage
Used in the "Tweak' and self linearization functions to enter the actual gas
concentrations.
See also