Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-63
NGA2000 Reference
ZERORNGS
Direction Bidirectional
Binding
None
Type Enumerated
Value range
SEPARATELY; TOGETHER
Units None
Description
Determines whether to zero each range separately or do them all at once.
Usage
If set to SEPARATELY, when a zero operation is performed, only the current range will
be zeroed. If set to TOGETHER, the analyzer will attempt to zero each range, one after
the other, assuming that the gas contains the concentration shown in each element of
ZEROGAS. If these are different, this will fail, unless CALCHKLIMITS is set to
DISABLED, in which case it will zero all ranges anyway. If the preamp gain has to be
changed between ranges, the analyzer will not zero any range for which it has to
change the gain. If it fails to zero any range, it will set CALRESULT such that any
failed ranges are uncalibrated. In this case it will leave the original zero factor
unchanged.
See also
ZERO
CALRANGES