Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-10
NGA2000 Reference
CALCHKLIMITS
Direction Bidirectional
Binding
CALCHKLIMITS
Type Enumerated
type
Value range
DISABLED; ENABLED
Units None
Description
Enables or disables calibration limit checking.
Usage
If enabled, will not allow a zero or span that requires more of a change (as a
percentage of the current range) than is shown in CALFPC. If disabled, will allow
zero or span to work no matter how much change in the calibration factors is made.
If the current calibration is known to be good, enabling this will prevent subsequent
calibration on grossly incorrect gases; if the current calibration is known to be bad for
some reason, disabling this will allow a successful zero or span no matter how much
of a change is required.
See also
CAL_ERR_MSG
CAL_VALIDITY
SPAN
ZERO
SPANGAS
ZEROGAS
CALDATE_S
Direction Output
Binding
None
Type String
Value range
31 characters
Units None
Description
Date of the last span
Usage
Shows the time and date of the last span. This requires that a control module be
used with the analyzer so that its internal clock is updated, or alternatively a PC can
be used to write into I_TIME.
See also
CALDATE_Z
CALDATE_Z
Direction Output
Binding
None
Type String
Value range
31 characters
Units None
Description
Date of the last zero
Usage
Shows the time and date of the last zero. This requires that a control module be
used with the analyzer so that its internal clock is updated, or alternatively a PC can
be used to write into I_TIME.
See also
CALDATE_S