NGA2000 Reference
TLIMB
Direction Bidirectional
Binding
None
Type
NDIR: 2-element floating point array, HFID: 4-element floating point array.
Value range
User selectable, depending on application
Units C
Description
Internal temperature measurement limits:
NDIR:
Detector temperature upper limit.
Detector temperature lower limit.
HFID:
Oven temperature upper limit.
Oven temperature lower limit.
Sample temperature upper limit.
Sample temperature lower limit.
Usage
Limits on the temperature measurements. If transgressed, the alarm system will issue
warning alarms.
See also
TLIMA
TRENDCONT
Direction Bidirectional
Binding
None
Type
2-element enumerated array.
Value range
A list of analyzer variables
Units None
Description
Lists possible analyzer variables whose 15 minute averages may be stored in the
internal storage buffer.
Usage
This variable shares data with TRENDVAR. TRENDVAR contains enumerations
whose phrases describe the variables whose names are referred to by the
enumerations in TRENDCONT. In the menu system, the user selects an element in
TRENDVAR and the system tells the software in the analyzer to store the data in the
variable pointed to by TRENCONT.
See also
TRENDVAR
TRENDTIME
Direction Bidirectional
Binding
None
Type Integer
Value range
0-240
Units Hours
Description
Length of time over which TRENDCONT and TRENDVAR are stored.
Usage
Not used. If implemented, will control the length of time over which the internal storage
will run. In version 2.2 or earlier, this is fixed at 24 hours.
See also
TRENDVAR