User guide
ELI
O
S
ECH001XN111-A4 – 07/2022
83
Handheld mode measurement
4.9 Audiometry
To build the curve you have to specify the patient answer with the “
Yes
” “
No
” buttons for each stimulation (activated
thanks to the “stim” button).
You can see the curve at any time by pressing the “
Graph
” button. Then you can
save
the curve by clicking on“
Save
”,
delete
it by exiting the window or
continue
the measure by clicking on one of the boxes in the summary table.
4.9.1.4
Weber test
The Weber test can detect if the patient presents a strong hearing lateralization. This allows to set at best the contra-
lateral masking noise power.
Patient setup
The Weber test is realized with the bone stimulator placed in the middle of the patient forehead.
Measurement
The bellow windows give access to the stimulation configuration.
The aim is to know for each frequency what is the threshold at which the patient hears on only one side.
For each stimulation, indicate if the patient hears the sound at the left, right or on both sides.
You can see the curve at any time by pressing the “
Graph
” button. Then you can
save
the curve by clicking on
“
Save
”,
delete
it by exiting the window or
continue
the measure with the “
Measure
” button.
4.9.2
High frequency audiometry
To make high frequency audiometry, it is mandatory to have a high frequency audiometric headset and have a license
for the high frequency module. The headset must be selection in the menu « headset » under the configuration window
(see
Erreur ! Source du renvoi introuvable.
).
To obtain more information about curve consulting, please refer to the paragraph
Erreur ! Source du renvoi
introuvable.
.
The saved data can be found in the “
Consultation
” patient menu.
Frequency selection
Start stimulation
Power setting
Patient answer
Curve consultation
It is extremely important to choose the correct headset model to ensure that the calibration is correctly ap-
ply when making measurement.
To obtain more information about curve consulting, please refer to the paragraph
Erreur ! Source du renvoi
introuvable.
.
The saved data can be found in the “
Consultation
” patient menu.
.