User guide
ELI
O
S
ECH001XN111-A4 – 07/2022
123
Measurement on ECHOSOFT
6.4 DP-gram
6.4.2
Description of the measurement window
1.
Graphic display area:
•
X-axis: frequency.
•
Y-axis: Power.
•
Green curve: effective signal power.
•
Black numbers: value in dB of the effective signal power.
•
Red curve: noise power.
2.
Table summarizing all the scanned frequencies:
•
Target power of L1.
•
Target power of L2.
•
Frequency F1.
•
Frequency F2.
•
Measured power of L1.
•
Measured power of L2.
•
Distortion product frequency.
•
Distortion product power.
•
Average level of noise.
•
Signal/noise ratio.
3.
Information about the selected validation protocol. For the "Screening" mode, in addition to the protocol used, a
pictogram is displayed representing the validation or not of the measurement.
4.
Note input.
5.
Options to print the measurement on paper or as a PDF file (for left and right printing on the same report, refer to
the paragraph below) and display information about the device and the test operator
Содержание ECHODIA ELIOS
Страница 100: ...User guide ELIOS ECH001XN111 A4 07 2022 99 Measurement on ECHOSOFT 6 1 Evoked potential modul ABR ECochG et VEMP...
Страница 118: ...User guide ELIOS ECH001XN111 A4 07 2022 117 Measurement on ECHOSOFT 6 3 Hydrops Shift OAE and DPMC...
Страница 137: ...User guide ELIOS ECH001XN111 A4 07 2022 136 Measurement on ECHOSOFT 6 6 Audiometry...