Introduction and test setup
3.2 ECochG
electrode (
Patient Reference
) is far less strict. This electrode can be placed on the forehead, on the temple or on
the chin
•
Connect the electrode in the middle of the forehead (
minus
) with the
Black
clip and the
Patient Reference
with
the
Green
clip
•
Roll and slightly crush the golden electrodes between your fingers, then insert it in the tested ear canals (
Red
tube
for the
Right
ear,
Blue
tube for the
Left
ear). If possible, after insertion of the golden electrode, apply a drop of saline
solution into the ear canal (this can improve the electrical conduction).
Position of the ear tip in the auditory canal
Incorrect insertion
Incorrect insertion
Be careful not to use too much physiological serum to avoid filling the patient’s ear canal.
The patient must be placed in comfortable way to avoid any excessive muscle tension.