Handheld mode measurement
4.7 DP-gram
4.7.3
Measurement consultation
This graph contains several items of information:
•
The green curve represents the power of the distortion product at the various selected frequencies in dB SPL.
•
The white number above a point is the power of the distortion product.
•
The vertical white numbers indicate the test frequency for each point.
•
The red curve represents the average noise level.
•
A reminder about the measurement conditions (power and number of simulations) is given at the bottom of the
window
•
On each point there is a validation (or not) mark.
•
In screening mode, there is an indication of the measurement validation or not.
4.7.3.1
Advanced analysis tools
ELI
O
S
includes a high-performance tool palette that giving you the opportunity of analyzing directly on the touch
screen (without any computer support), all the data you have collected.
Click on one of the points of the curve. A window including a data analysis table appears. It contains several items of
information concerning the investigated signal.
Refer to paragraph
for more details about patient management.
Signal level in dB SPL
Stimulation power
Frequencies in Hz
dB SPL scale
PASS/REFER indicator
(only in screening mode)
Signal level
validation indicator
Validation protocol
Noise level
The column in green represents the distortion products investigated in the previous phase graph. This value
can be modified. Refer to paragraph
Erreur ! Source du renvoi introuvable.
.
2
∗
F
1
−
F
2
Main distortion product
F
1
Stimulation frequency F1
F
2
Stimulation frequency F2
2
∗
F
2
−
F
1
Secondary distortion product
FREQ
Frequency in Hz
POWER
Power in dB
PHASE
Phase shift in degrees
S/N
𝑆𝑖𝑔𝑛𝑎𝑙
𝑁𝑜𝑖𝑠𝑒
ratio in dB