Handheld mode measurement
4.8 TEOAE
5dB to strong). This may be due to the wrong positioning of the probe, more precisely if the probe is to deep in the
ear canal. It is advised to click on
“No”
, reposition the probe and restart the measure. However, if you are sure of
the probe’s positioning, it is possible to continue by clicking
“Yes”
.
Once the calibration is done, the measurement window appears :
•
On the left side:
o
The rejection indicator light notifies you when the rejection threshold is reached. This phenomenon indicates
that the patient moves or, in common way, that there is too much noise. When the noise will go down, the
measure will automatically resume
o
Indicator of the stimulation quality: in a good measurement condition, the gauge should be green and half
filled. If the gauge moves or change of color, this is because of a bad positioning of the probe, or a clogged
probe.
•
The graph show :
o
In the left part, the shape of the stimulation
o
In the central part, the two buffer (A and B) that are currently recorded.
•
At the bottom of the screen:
o
The number of clicks, which informs you of the progress of the measurement, the sum of the two buffers (A
and B) must reach the number of stimulations entered in the configuration window.
o
The correlation between the A and B buffers
o
Number of rejections for each buffer
o
Detection analysis, which works only in detection mode. It shows how many frequencies have been validated
or how many statistical criteria have been satisfied, depending on the validation mode chosen. (See subsection
The "
Stop
" button stops the measurement prematurely but the data is not lost, you will be able to consult the meas-
urement. You then have the choice of saving the data by clicking on "
Save
" or deleting it by exiting from the window
using the back cross.
4.8.1.2
Prerequisites for exploitation of the measure
If the click shape doesn't look like the above illustration (sinusoid amortized on a few oscillations) check the probe
position in the ear then restart the measurement.
Rejection indicator
Stimulation curve
Stimulation quality
indicator
Number of click done for
each buffer
TEOAE response
area
Buffer A curve
Correlation between buffer
Number of rejections
per buffer
Validation analysis
Buffer B curve
Содержание ECHODIA ELIOS
Страница 100: ...User guide ELIOS ECH001XN111 A4 07 2022 99 Measurement on ECHOSOFT 6 1 Evoked potential modul ABR ECochG et VEMP...
Страница 118: ...User guide ELIOS ECH001XN111 A4 07 2022 117 Measurement on ECHOSOFT 6 3 Hydrops Shift OAE and DPMC...
Страница 137: ...User guide ELIOS ECH001XN111 A4 07 2022 136 Measurement on ECHOSOFT 6 6 Audiometry...