User guide
ELI
O
S
ECH001XN111-A4 – 07/2022
78
Handheld mode measurement
4.8 TEOAE
4.8.2
Screening mode
In the "
Screening
" mode it is not possible to adjust the
stimulation rate (fixed at 80Hz). In this mode the device
stops the measurement when the validation criteria are
satisfied. On the other hand, if the maximum test dura-
tion is reached, the measurement is stopped and the de-
vice indicates a non-conclusive test.
The validation criteria and the maximum test duration can be modified in the advanced setting of the TEOAE (see
). In this menu, it is also possible to modify the intensity used in
« Screening »
mode. The intensity
setting and the access to this menu directly from the measurement window can be unlocked in the
« Config OAE »
(see
subsection
Erreur ! Source du renvoi introuvable.
).
At the end of the measurement, a popup shows if the test is valid or not.
End of a valid test
End of a test that can’t be validated
If you are not familiar with the settings and how modifying them can affect test results, do not attempt to
change them. A bad setting can make the next measurements unusable and irrelevant.
This screening mode is more suited to newborns.
For more details about the curve consulting options, please refer to the paragraph
Erreur ! Source du renvoi
introuvable.
.
The saved data can be consulted in the patient's "
Consultation
" menu.