Introduction and test setup
3.3 VEMP
Measurement with DD45 headphone
DD45 headphone
3.3.1
Patient setup
•
Connect the
Red
clip tool to the
Red
electrophysiological cable and the
Blue
clip to the
Blue
electrophysiologi-
cal cable
.
•
Connect the electrophysiological cable
to the
ECHO-DIF
. Connect the
ECHO-DIF
Mini-DIN on the
AUX
con-
nector.
•
For measurement with the insert earphone, put the tip on the
left
and
right
earphones. Then, connect the ear-
phone Mini-DIN to the
«Audio»
connector of the
ELI
O
S
unit.
•
Roll the foam ear tips on the
Red
stimulator between your fingers then insert it in the
right
ear. Then insert the ear
tips plug on
Blue
acoustic stimulator in
left
ear. The
Right
/
Left
switch is automatically done.
Position of the ear tip in the auditory canal
Incorrect insertion
Incorrect insertion
•
To make measurement with
DD45 headset
, connect the headset cable on the Jack plug of the
ELI
O
S
(indicated
with the headset icon).
•
Clean the surface of the skin where the electrodes will be attached with abrasive gel. This decreases the impedance
of the skin. Depending on the contact being used, it may be necessary to remove the deposit with a cleaning agent
(such as alcohol).
Using an otoscope, make sure that the ear canal is not obstructed by ear-wax.
This operation must be carried out by a qualified person.
These instructions must be adapted depending of the ear(s) tested, in every case the
red
color corresponds
to the
right
ear, the
blue
color to the
left
ear.