User guide
ELI
O
S
ECH001XN111-A4 – 07/2022
122
Measurement on ECHOSOFT
6.4 DP-gram
6.4
DP-gram
to obtain the necessary instructions about the needed equipment and the setup.
6.4.1
Description of the measurement window
For details about presentation and analysis of the curves, see section
Erreur ! Source du renvoi introuvable.
.
1.
L2 power selection in dB SPL (The difference between the L1 and L2 values can be set in the advanced settings of
the device – see subsection
).
2.
Select the number of average (number of stimulation) for one point (this setting is not available in «Screening»
mode).
3.
Rejection setting: the value corresponds to the acoustic noise level in dB SPL above which the acquisition will be
rejected
.
.
4.
Select measurement frequencies.
5.
«Screening» mode
.
In this mode, the device switches to the next frequency when the validation conditions are
reached or if the maximum test time has been reached. After testing all the selected frequencies, the instrument
stops the measurement and indicates whether the test is valid or not, depending on the number of frequencies at
which the distortion product (DP) was detected.
6.
Display of an explicative image for the connection of the device and the patient (as in
Erreur ! Source du renvoi
introuvable.
)
7.
Select the ear side.
8.
Display both ear side in the same window.
9.
Start a new measurement.
10.
Allows you to move on to the next stimulation frequency when it is not going well (excessive patient tension,
excessive noise in the measurement room...).
11.
Save the current measurement or create a new one.
12.
Information on measurement progress (number of acquisitions, acquisition speed, number of rejected acquisitions)
13.
Information about the selected validation protocol. For the "Screening" mode, in addition to the protocol used, a
pictogram is displayed representing the validation or not of the measurement. The validation conditions, as well as
the maximum duration of the test, can be modified in the advanced settings of the DPgram. (see subsection