Measurement on ECHOSOFT
6.1 Evoked potential modul (ABR, ECochG et VEMP)
6.1.5
ABR screening mode
Among the tests of the evoked potentials module, the screening mode is only available for the ABR. This measure-
ment mode is intended for newborns. The configuration options are limited, but it ensures a fully automated measurement
and diagnosis.
1.
Power: by convention, the screening test for new-
borns should be carried out between 35 and 40 dB.
However, it is possible to modify this value, but
above 50 dB, the device displays a message indi-
cating that the measurement no longer corresponds
to a screening test.
2.
Choice of stimulation and collection side: If both
ears are selected, the right ear will be tested first,
then the left.
3.
Replace the Click stimulus by a Chirp stimulus
(the Chirp stimulus is only used in screening)
4.
Must be selected when using insert stimulators with a tube.
5.
Choose between an installation with only 3 electrodes instead of the classic installation with 4 electrodes (see
paragraph
Erreur ! Source du renvoi introuvable.
). Rather than using a red and blue electrode on their respec-
tive mastoids, here it is possible to use only the red electrode placed on the neck of the newborn.
Simplify setup with 3 electrodes
6.
Start and stop measurement.
Start of the recording
Time in µs
Time in µs
150µs clic
Burst at 1000Hz 3/3/3