ELI
O
S
ECH001XN111-A4 – 07/2022
33
Introduction and test setup
3.4 ASSR
3.4.1
Patient setup
•
Connect the
Red
clip tool to the
Red
electrophysiological cable and the
Blue
clip to the
Blue
electrophysiologi-
cal cable
.
•
Connect the electrophysiological cable
to the
ECHO-DIF
. Connect the
ECHO-DIF
Mini-DIN on the
AUX
con-
nector.
•
For measurement with the insert earphone, put the tip on the
left
and
right
earphones. Then, connect the ear-
phone Mini-DIN to the
«Audio»
connector of the
ELI
O
S
unit.
If you have the « pediatric kit », you can use the acoustic tubes with OAE ear tips connected on the insert ear-
phone instead of the foam ear tip (the
«tube»
option must be activated on the software).
•
To make measurement with
DD45 headset
, connect the headset cable on the Jack plug of the
ELI
O
S
(indicated
with the headset icon).
•
Clean the surface of the skin where the electrodes will be attached with abrasive gel. This decreases the impedance
of the skin. Depending on the contact being used, it may be necessary to remove the deposit with a cleaning agent
(such as alcohol).
•
Attach an electrode (
minus
) in the middle of the forehead, just below the hairline. The positioning of the other
electrode (
Patient Reference
) is far less strict. This electrode can be placed on the forehead, on the temple or on
the chin.
•
The electrodes
V+
and
V+
must be attached behind the ear to be tested (on the mastoid)
Using an otoscope, make sure that the ear canal is not obstructed by ear-wax.
This operation must be carried out by a qualified person.
These instructions must be adapted depending of the ear(s) tested, in every case the
red
color corresponds
to the
right
ear, the
blue
color to the
left
ear.