User guide
ELI
O
S
ECH001XN111-A4 – 07/2022
23
Introduction and test setup
3.1 ABR
electrode (
Patient Reference
) is far less strict. This electrode can be placed on the forehead, on the temple or on
the chin.
•
The electrodes
V+
and
V+
must be attached behind the ear to be tested (on the mastoid)
•
Connect the electrode in the middle of the forehead (
minus
) with the
Black
clip and the
Patient Reference
with the
Green
clip. The
Red
clip must be connected to the electrodes placed behind the
Right
ear and the
Blue
one behind
the
Left
ear. The
Right
/
Left
switch is automatically done.
•
Roll the foam ear tips on the
Red
stimulator between your fingers then insert it in the
right
ear. Then insert the ear
tips plug on
Blue
acoustic stimulator in
left
ear. The
Right
/
Left
switch is automatically done.
Position of the ear tip in the auditory canal
Incorrect insertion
Correct insertion
The patient must be placed in comfortable way to avoid any excessive muscle tension.
Содержание ECHODIA ELIOS
Страница 100: ...User guide ELIOS ECH001XN111 A4 07 2022 99 Measurement on ECHOSOFT 6 1 Evoked potential modul ABR ECochG et VEMP...
Страница 118: ...User guide ELIOS ECH001XN111 A4 07 2022 117 Measurement on ECHOSOFT 6 3 Hydrops Shift OAE and DPMC...
Страница 137: ...User guide ELIOS ECH001XN111 A4 07 2022 136 Measurement on ECHOSOFT 6 6 Audiometry...