Introduction and test setup
3.3 VEMP
3.3.1.1
cVEMP
•
Attach an electrode
Patient Reference
in the middle of the forehead, just below the hairline. The other electrode,
(
minus
) is placed on the
sternum
.
•
The electrodes
V+
and
V+
must be placed on the sternocleidomastoid muscles.
•
Connect the electrode in the middle of the forehead (
ref patient
) with the
green
clip and the
V-
with the
black
clip. The
Red
clip must be connected to the electrodes placed on the
Right
sternocleidomastoid muscle and the
Blue
one the
Left
s
ternocleidomastoid muscle. The
Right
/
Left
switch is automatically done.
Position
For the
cVEMP
measurement, the patient’s position is very important because it directly affects the measurement quality.
Indeed, to correctly collect data, the
s
ternocleidomastoid muscle must be contracted enough. The best way for that is to lie
down the patient and ask him to keep his head elevated in the opposite direction of the stimulation.
To make measurement on both ears easier, it is recommended to equip one ear at a time because the meas-
urement position is hard to hold when the electrodes and clips are on both sides.