Introduction and test setup
3.4 ASSR
•
Connect the electrode in the middle of the forehead (
minus
) with the
Black
clip and the
Patient Reference
with the
Green
clip. The
Red
clip must be connected to the electrodes placed behind the
Right
ear and the
Blue
one behind
the
Left
ear. The
Right
/
Left
switch is automatically done.
•
Roll the foam ear tips on the
Red
stimulator between your fingers then insert it in the
right
ear. Then insert the ear
tips plug on
Blue
acoustic stimulator in
left
ear. The
Right
/
Left
switch is automatically done.
Position of the ear tip in the auditory canal
Incorrect insertion
Incorrect insertion
As the test can last up to 40 minutes in the case of a threshold search, the patient must be placed in comforta-
ble way to avoid any excessive muscle tension and discomfort.