User guide
ELI
O
S
ECH001XN111-A4 – 07/2022
143
Technical specifications
8.1 General technical characteristic of the device
8.1.1
Test parameters :
Measurement
Characteristic
Shift-OAE (DPOAE)
- Acoustic stimulation: from 1 kHz to 3 kHz
- Digital resolution 16 bits @ 32kHz
- Sound intensity: 50 to 75 dB SPL
DPMC
- Acoustic stimulation: 900Hz to 1100Hz
- Specific earplug
- Digital resolution 16 bits @ 32kHz
- Impedance test
- Configurable rejection
- Sound intensity: 50 to 90 dB SPL
DP-gram
- Acoustic stimulation: 1kHz to 5kHz
- Digital resolution 16 bits @ 32kHz
- Sound intensity: 50 to 75 dB SPL
TEOAE
- 25 clicks per second
- Alternating clicks by buffer of 4
- Digital resolution 16 bits @ 32KHz
- Sound intensity: 40 to 95 dB SPL
ABR
- Up to 50 clicks per second
- Alternating clicks
- Digital resolution 16 bits @ 32KHz
- Impedance test
- Measurement window from 10 to 25 ms
- Sound intensity: 0 to 95 dB HL
ASSR
-AM2 stimulation
-Carrier frequencies at 500Hz, 1000Hz, 2000Hz, 4000Hz
-Modulation at 40Hz or 80Hz
-Digital resolution 16 bits @ 32KHz
- Impedance test
- Sound intensity: 10 to 95 dB HL
ECochG
- Up to 50 clicks per second
- Alternating clicks
- Digital resolution 16 bits @ 32KHz
- Impedance test
- Measurement window from 10 to 25 ms
- Sound intensity: 0 to 95 dB HL
VEMP
- Up to 50 clicks per second
- Alternating clicks
- Digital resolution 16 bits @ 32KHz
- Impedance test
- Measurement window up to 60 ms
- Sound intensity: 0 to 105 dB HL
Pure tone audiometry
- Sound intensity AC: from -10 to 110 dB HL
- Sound intensity BC: from -10 to 80 dB HL
- Available intensity step: 5 dB
- Acoustic Stimulation: 125Hz to 8kHz
(up to 16kHz with HF module)
-Narrow band masking noise: 1/3 octave
- Manual operation
- Automatic operation
Speech audiometry
- Sound intensity: from -10 to 110 dB HL
- Automatic list selection