Service Mode Savina Version 2.0
V_TEMP_AWT
1023
5.00V
(without external AWT sensor).
Measurement signal of AWT sensor.
The measurement signal is processed
on the Control PCB (tempera
hybrid).
V_TEMP_AWT
0716 ... 0818 3.50 V ... 4.00 V (with external AWT sensor, at a
temperature).
V_FLOW
0004 ... 0012 0.02 V ... 0.06 V (with flow exp. = 0). Flow signal o
expiratory flow sensor S2.
V_FLOW_NULL
0822 ... 0831 4.02 V ... 4.06 V (with flow exp. = 0). Signal of ex
flow sensor S2.
V_FLOW_INOP
0102 ... 0246 0.50 V ... 1.20 V Signal of expiratory flow senso