Error List Savina Version 2.0
40.0000
Power switch defective.
−
Check power switch and cable
harness.
−
Check connection to connector X12 on
Control PCB.
−
Replace power switch.
−
See also
”
Unit cannot be switched off
in chapter
”
Repair Information/Error
Events
”
.
40.0010
DC internal defective.
−
Savina does not recognize the internal
battery due to complete discharge,
therefore pre-charge the internal
battery using an external charger (see
also
”
Warning message "!!! No int.
battery"
”
in chapter
”
Repair
Information/Error Events
”
).
−
Install new internal battery.