B
C
3.2.3
Patient leakage current test
Connect airway temperature sensor (1) to the
socket of the airway temperature sensor of
Savina and to the tester.
Enter serial number of the airway temperature
sensor:
Normal condition (N.C.):
I
≤
100
µ
A
Single fault condition (S.F.C.):
Neutral conductor interrupted:
I
≤
500
µ
A
In the following steps the test is repeated, but
with the power plug turned over. This
condition can be created internally with some
types of test devices.
Normal condition (N.C.):
I
≤
100
µ
A
Single fault condition (S.F.C.):
Neutral conductor interrupted:
I
≤
500
µ
A
3.2.4
Multiple socket outlet on trolley (if fitted)
3.2.4.1
Compare fuses with rated values on the
socket-outlet.
3.2.4.2
Protective earth conductor test
Check all sockets.
Test value R
≤
0.2 ohms
Tester
Savina
Mains voltage
L/N
N/L
PE
mA
PS
Patient leakage current test
Patient leakage current test
(1)
P
P
P
P
P
Z
P