![Dräger Savina Скачать руководство пользователя страница 137](http://html.mh-extra.com/html/drager/savina/savina_service-manual_2526836137.webp)
F
o
r
i
n
t
e
r
n
a
l
u
s
e
o
n
l
y
.
C
o
p
y
r
i
g
h
t
r
e
s
e
r
v
e
d
.
R
5
6
6
4
9
0
0
F
L
1
.
f
m
2
7
.
0
8
.
0
1
A
B
C
Dr
ä
ger Medizintechnik GmbH
Error List Savina Version 2.0
42.0000
Actual speed of blower is too slow.
−
In new device it may occur that the
blower is not fully run in which could
lead to this sporadic error.
−
Replace the blower.
42.0001
Actual speed of blower is too fast.
−
In new device it may occur that the
blower is not fully run in which could
lead to this sporadic error.
−
Replace the blower.
42.0002
Deviation between set and actual
speed.
−
In new device it may occur that the
blower is not fully run in which could
lead to this sporadic error.
−
Replace the blower.
42.0100
Micro-filter not fitted (switch is open).
−
Micro-filter not fitted or not fitted
correctly.
−
Switch behind the micro-filter is
42.0110
Resistance of the micro-filter is too
high (not used for the time being).
−