34
A
B
C
4.6
Testing "Emergency respiratory valve D2
(emergency air valve)"
Test set-up:
Switch off the unit. Through the test pressure
regulator with injector set a flow of
60 L/min ±3 L/min.
Test value:
The pressure at the Y-piece must be between
—3 mbar and —6 mbar.
4.7
O2 supply pressure monitoring
Switch on the unit.
Disconnect the O2 supply.
Set the O2 concentration to 50%.
Audible alarm sounds and display message
"!!! No O2" appears.
Restore the O2 supply.
Savina
Pressure gauge —30 to 120 mbar
Flowmeter 10 to 120 L/min
Length of hoses up to Y-piece:
1.2 m each
P
F