Dr
ä
ger Medizintechnik GmbH
Error List Savina Version 2.0
40.1030
Power pack is no longer electrically
safe.
−
Replace power pack.
−
Voltage of internal battery is greater
than 30 V. Replace internal battery
−
External DC voltage is greater than
15.8 V. The problem could be due to
two completely discharged external
batteries, that is the power pack only
detects one battery. Disconnect
external battery and recharge or
disconnect and reconnect batteries.
−
External DC voltage is greater than
31.8 V. Replace external batterie
−
Check internal and external batteries.
Replace power pack, if necessary
40.1100
Communication (1) between power
pack and Control PCB is faulty.
−
Could be due to a log error after fast
change of status. Power Savina from
mains supply.
−
Replace power pack.