Dr
ä
ger Medizintechnik GmbH
Repair Procedure Savina Version 2.0
Voltage test points on the Control PCB
Voltage
Pin assignment, connector X1
Test point
+5 V
2/4
T43
DGND
1/3
T44
-15 V
11/12
T41
+15 V
8/9
T16
AGND
7/10
T39
+24 V
6
T625
AGND
5
T626
Battery
The battery voltage must be measured directly at the battery (see Mainten
Procedure, chapter "
Replacing the internal rechargeable batteries
"). It can also be
measured between the battery fuse and the external power supply connection (see
also "
"Starting aid" for charging of internal battery
").
X
2
X
1
X
6
4
5
X
9
X648
X652
X401 X402
X403
s
o
r
B
a
s
s
e
m
b
l
y
D
i
s
p
l
a
y
park
position
i
n
t
E
x
p
i
r
a
t
o
r
y
v
a
l
v
e
P
o
w
e
r
p
a
c
k
D
C
o
u
t
p
u
t
s
P
o
w
e
r
p
a
c
k
c
h
a
r
g
i
n
g
m
a
n
a
g
e
m
e
n
P
o
w
e
r
p
a
c
k
I
I
I
V
a
l
v
e
d
r
i
v
e
C
o
n
t
r
o
l
D40
X15
X635
X625
D32
6 4 2
X2
X1
X645
X23
X9
T626
T625
T44
T43
T41
T16
T39
X20