Service Mode Savina Version 2.0
Test step 13 is used to test the nebulizer valve (V5). The valve can be switched on or off. When the
valve is switched on, O2 flows out of the nebulizer socket (8 L/min +5 L/min). The nebulize
powered with O2.
"Insp.cal valve" (V6.1)
"Exp.cal valve" V6.2
"Flowinsp"
"Pawinsp"
"Pawex
on
off
10 L/min
0 mbar
±0.3 mbar
approx
off
off
10 L/min
approx.
40 mbar approx.
4
off
on
10 L/min
approx.
40 mbar 0 mbar
±0.3 mb
on
on
10 L/min
0 mbar
±0.3 mbar
0 mba
±0.3 mb
Connect (short-circuit) the inspiratory socket and the expiratory socket with a v
hose to perform this test step.