ger Medizintechnik GmbH
Software Download Savina Version 2.0
•
Under the "Flash programming/Master" menu enter the path of the "8414344.dld" file.
•
Under the "Flash programming/Front" menu enter the path of the "8414345.dld" file.
•
Switch off Savina.
•
Cut all the power to the Savina:
−
Pull the power plug.
−
Remove the fuse for the external battery (see also "
Fig.17:
in the maintenance procedure"
section).
−
Cut any power from an external battery or from the on-board power supply.
•
Open the Savina operator control panel (see also "
Opening the operator control panel
"Repair procedure" section).
•
Plug a jumper onto pins 5 and 6 of connector X14 and another one onto pins 5 and 6
connector X15 on the Control PCB (see also "
View, Control PCB
" in the "Schematics and
Diagrams" section).