Error List Savina Version 2.0
41.0040
Operating v5 V of the ambient
pressure sensors is too low.
−
Operating v15 V prese
"Repair Procedure
Measuring the
supply voltages
")?
−
Operating v15 V for O2 s
pressure sensor on O2 Valve PCB
present (test point T2(+) and T9
(AGND1) on the O2 Valve PCB,
also chapter "Schematics and
Diagrams
O2 Valve PCB
")?
−
Operating v5 V for O2 am
pressure sensors on O2 Valve PCB
present (test point T7(+) and T9
(AGND1) on the O2 Valve PCB,
also chapter "Schematics and
Diagrams
O2 Valve PCB
")?
−
Replace the O2 Valve PCB.