Dr
ä
ger Medizintechnik GmbH
Service Mode Savina Version 2.0
Test Step 15
Setting the calibration classes for the bypass valve (V1) and of the expiratory
valve (V3).
Test Step 16
Displaying the internal device temperature.
Test Step 17
Displaying the internal, analog measured values of the master processor.
Test Step 18
—
see
Test Step 17
—
Test Step 19
Displaying the external, analog measured values of the master processor
Test Step 20
Displaying the internal, analog measured values of the front processor.
Test Step 21
—
see
Test Step 20
—
Test Step 22
Displaying and calibrating the offset and calibration voltage of the electronics for
the O
2
sensors.
Test Step 23
Calibrating the offset voltage of the inspiratory and expiratory pressure sensors.
Test Step 24
Displaying the calibration data of the O
2
sensors.
Test Step 25
—
Up to SW 1.01, not assigned
—
as of SW 1.10, display of actual status of
display, 7-segment display, and LEDs.
Test Step 26
—
Up to SW 1.01, not assigned
—
as of SW 1.10, setting factory default values.
Test Step 27
Entering the ambient pressure.