Error List Savina Version 2.0
01.0306
Data retention (EEPROM), checksum
error: Spirolog sensor (not used for the
time being).
−
01.0307
Data retention (EEPROM), checksum
error: Offset of O2 sensor 1.
−
Carry out Service Mode, test 22.
−
Replace Control PCB.
01.0308
Data retention (EEPROM), checksum
error: Calibration of O2 sensor 1.
−
Switch the unit off/on. O2 sensor 1
calibrated after power-on.
−
Replace Control PCB.
01.0309
Data retention (EEPROM), checksum
error: Offset of O2 sensor 2.
−
Carry out Service Mode, test 22.
−
Replace Control PCB.
01.0310
Data retention (EEPROM), checksum
error: Calibration of O2 sensor 2.
−
Calibrate O2 sensor 2 (menu
"O2 calib.").
−
Replace Control PCB.
01.0311
Data retention (EEPROM), checksum
error: Calibration of O2 supply pressure
sensor S5 , ambient pressure sensors
S6 and S7.
−
Carry out Service Mode, test 27
(calibration of ambient pressure
sensors).