C
Dr
ä
ger Medizintechnik GmbH
Function Description Savina Version 2.0
The power pack delivers the following output voltages:
−
+5 V
−
—
15 V
−
+15 V
−
+24 V
−
+48 V
The output voltages are short-circuit-proof and stable at no-load.
The output voltages are generated according to the following priority, dependent on the
voltages:
Input voltage
Priority
Action
AC voltage
1
Charge external and internal batteries, and maintain the c
External on-board
power supply
2
Charge internal batteries, and maintain the charge.
External
battery/batteries
3
Maintain the charge of the internal batteries.
Internal batteries
4
—