Dr
ä
ger Medizintechnik GmbH
Error List Savina Version 2.0
00.2020
Device reset triggered: SW error.
−
Report error to DS L
ü
beck.
00.2100
Device reset triggered: SW error.
−
Report error to DS L
ü
beck.
00.2110
Device reset triggered: SW error.
−
Report error to DS L
ü
beck.
01.0000
ROM test: Checksum procedure test
not OK.
−
Carry out SW download using current
SW.
−
Replace Control PCB.
01.0100
ROM test: Error found during
checksum start test.
−
Carry out SW download using current
SW.
−
Replace Control PCB.
01.0101
ROM test: Error found during current
checksum test.
−
Carry out SW download using current
SW.
−
Replace Control PCB.
01.0200
Data retention: Read error.
−
Replace Control PCB.
01.0201
Data retention: Write error.
−
Replace Control PCB.