Alarm Messages Savina Version 2.0
1.
The oxygen control cannot reach the correct FiO2 set.
1.
The measured value FiO2(t) exceeds the limit value FiO2 high. This error detection i
provided with an error counter.
1.
After a cold start and 2 minutes afterwards.
2.
During standby and 2 minutes afterwards.
3.
When O2 monitoring is off and 2 minutes afterwards.
4.
During calibration of the O2 sensor and 2 minutes afterwards.
5.
After adjusting the FIO2
target value
and 2 minutes afterwards.
1.
Generation of event "FiO2 high" (warning "!!! FiO2 high").